Model Number G141 |
Device Problem
Impedance Problem (2950)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/17/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) detected out-of-range impedance measurements on the right ventricular (rv) lead.Technical services was consulted and provided lead troubleshooting recommendations.No adverse patient effects were reported.Of note, the rv lead is a non-boston scientific product.The crt-d remains in service.
|
|
Event Description
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) detected out-of-range impedance measurements on the right ventricular (rv) lead.Technical services was consulted and provided lead troubleshooting recommendations.No adverse patient effects were reported.Of note, the rv lead is a non-boston scientific product.The crt-d remains in service.
|
|
Manufacturer Narrative
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined this device exhibited intermittent impedance measurements with no conclusive evidence of a malfunction or inadequate lead-to-device connection; please refer to the event description for more information regarding the specific circumstances of this event.Investigation has determined this type of event is likely the result of an intermittent high impedance condition associated with the device spring contact and lead terminal ring.A design enhancement was implemented in 2020 to stabilize the electrical connection between the spring contact and the lead terminal.
|
|