ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY ASSURA CRT-D RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
|
Back to Search Results |
|
Model Number CD3357-40Q |
Device Problems
Failure to Capture (1081); Failure to Sense (1559)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/16/2023 |
Event Type
malfunction
|
Event Description
|
Related manufacturer reference number: 2017865-2023-02857.It was reported that the patient presented in clinic for follow up.Upon device interrogation, the right ventricular (rv) lead exhibited under-sensing, loss of sensing and capture.The lead was capped and replaced on (b)(6) 2023.During the same procedure, after the right ventricular lead was replaced, the implantable cardioverter defibrillator exhibited loss of sensing and capture.The device was explanted and replaced.The patient condition was stable.
|
|
Manufacturer Narrative
|
The reported field event of under sensing and failure to capture could not be reproduced during analysis.Telemetry, impedance, sensing, pacing, and high voltage (hv) output functions of the device were tested and found to be normal.
|
|