Model Number CD3357-40Q
Device Problems Failure to Capture (1081); Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/16/2023
Event Type  malfunction  
Event Description
Related manufacturer reference number: 2017865-2023-02857.It was reported that the patient presented in clinic for follow up.Upon device interrogation, the right ventricular (rv) lead exhibited under-sensing, loss of sensing and capture.The lead was capped and replaced on (b)(6) 2023.During the same procedure, after the right ventricular lead was replaced, the implantable cardioverter defibrillator exhibited loss of sensing and capture.The device was explanted and replaced.The patient condition was stable.
 
Manufacturer Narrative
The reported field event of under sensing and failure to capture could not be reproduced during analysis.Telemetry, impedance, sensing, pacing, and high voltage (hv) output functions of the device were tested and found to be normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA CRT-D RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16233222
MDR Text Key308014528
Report Number2017865-2023-02858
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508216
UDI-Public05414734508216
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2019
Device Model NumberCD3357-40Q
Device Catalogue NumberCD3357-40Q
Device Lot NumberA000042601
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/10/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
QUICKFLEXMICRO; TENDRIL
Patient Age34 YR
Patient SexMale
-
-