C.R. BARD, INC. (BASD) -3006260740 POWERPORT ISP MRI 6CF INT W SP, ATT, SL; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 8806060 |
Device Problems
Suction Problem (2170); Material Twisted/Bent (2981); Migration (4003)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 12/30/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
As the lot number for the device was provided, a review of the device history record is currently being performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiration date: 08/2023).
|
|
Event Description
|
It was reported that approximately five months post port placement, the middle of the line allegedly twisted and created a loop upwards into the jugular vein.It was further reported that the port could not aspirate blood.Reportedly, patient's line was migrated.The patient had no additional intervention and the port had been removed.The patient's current status is unknown.
|
|
Manufacturer Narrative
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failures as no objective evidence was provided for review.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 08/2023).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that approximately five months post port placement, the middle of the line allegedly twisted and created a loop upwards into the jugular vein.It was further reported that the port could not aspirate blood.Reportedly, patient's line was migrated.The patient had no additional intervention and the port had been removed.The patient's current status is unknown.
|
|
Search Alerts/Recalls
|
|
|