• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 X-PORT ISP IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 X-PORT ISP IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number 7707540
Device Problems Fracture (1260); Material Separation (1562); Migration (4003)
Patient Problems Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/31/2022
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The device has not been returned to the manufacturer for evaluation.However, an image was provided for review.The investigation of the reported event is currently underway.Expiry date: 09/2024.Device not returned.
 
Event Description
It was reported that approximately two months from post port placement, the catheter was allegedly fractured.It was further reported that the catheter was separated and migrated to right atrium of the heart.The current status of the patient is unknown.
 
Event Description
It was reported that two months from post port placement, the catheter was allegedly fractured.It was further reported that the catheter was separated and migrated to right atrium of the heart.The current status of the patient is unknown.
 
Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.However, one image was provided for review.The investigation is confirmed for the reported catheter fracture, material separation and migration issues as in the image the catheter is in the internal jugular extending into the right atrium.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 09/2024).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-PORT ISP IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key16251608
MDR Text Key308507436
Report Number3006260740-2023-00166
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741027499
UDI-Public(01)00801741027499
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K022983
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/27/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7707540
Device Catalogue Number7707540
Device Lot NumberREGQ2849
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/11/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age55 YR
Patient SexFemale
Patient Weight52 KG
-
-