MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 8637-20 |
Device Problems
Human-Device Interface Problem (2949); Inappropriate or Unexpected Reset (2959)
|
Patient Problems
Pain (1994); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/20/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a healthcare provider (hcp) via a manufacturer representative (rep) regarding an implanted pump system for spinal pain indications.The pump was used to deliver unknown morphine 1000mcg/ml.It was reported that, when the pump was read on (b)(6) 2023, the hcp noticed codes "101" and "87", indicating that the pump reset to minimum rate and was in safe mode.The rep saw the same errors.The pump logs showed an update when the pump was implanted at (b)(6) 2023 at around 9am and the reset at around 12pm on the same date.No other events were logged.It was confirmed that the patient did not have magnetic resonance imaging (mri) and that the patient was not around any significant electromagnetic interference (emi) at that time.It was initially stated that the alarm was not heard until around the time that it was read on (b)(6) 2023, but it was later confirmed that the patient "heard something" but did not know that it was coming from their pump.On (b)(6) 2023, the initial pump update would not start and it immediately stated that pending data was invalid.The rep silenced the alarm and added the personal therapy manager (ptm) data, as the ptm was disabled since the patient was not currently using the ptm.The update was then successful.Originally, the hcp was planning to change the concentration to 2000mcg/ml and increase the dose from 28mcg/day to 96 mcg/day.It was confirmed that the hcp was comfortable with increasing the dose to 96mcg for the bridge bolus and simple continuous dose after the bridge bolus.The second update with a concentration change, bridge bolus, and ptm disabling was successful.No patient symptoms or complications were reported.Troubleshooting resolved the reported issue.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from a consumer via a company representative (rep) and it was reported the patient thought she heard her pump alarm twice, but not continuously and reported patient had increased pain.The patient's pain was a lot higher again and pain was at a 9.It was noted the patient came in for an appointment on the date of this report.The rep connected to the pump, and nothing showed up in the logs.The rep also performed the alarm test and patient said she could not hear them.The rep confirmed the patient had mistaken previous sounds for pump alarming.No update or changes were made.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from the company representative reported that the cause of the increased pain was unidentified, but it has been managed/resolved at this time.Patient has had a dose increase and the healthcare provider planned to give them a personal therapy manager (ptm) at their next appointment later this week.There were no other concerns at this time.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from the company representative (rep) regarding the patient and healthcare provider information.There were no external/environmental/patient factors that may have caused or contributed to the pump reset and safe rate.The cause of pump resetting, and safe rate was unknown.The patient did not hear the alarm because her hearing aids was not on but would hear the alarm if their hearing was on.She thought the noise was part of her apple watch.At her appointment, both pump alarms were played for the patient and her daughter.The patient did not hear the sounds during examples, and the daughter was able to determine both and know the difference between the meaning and frequency of each.
|
|