NELLCOR PURITAN BENNETT IRELAND NELLCOR; OXIMETER Back to Search Results
Model Number PM1000N
Device Problems Device Alarm System (1012); Failure to Run on Battery (1466); Failure to Power Up (1476); Failure to Analyze Signal (1539); Loss of Data (2903); Noise, Audible (3273); Delayed Program or Algorithm Execution (4034)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/13/2023
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during use on an adult ventilated patient that receives round the clock care and was monitored during the night , the unit has been intermittently malfunctioning with gaps in trend data, alarms, and shutting itself down.The power and the battery were verified and restarted the unit for troubleshooting. the unit was used with normal parameters and the monitor was not modified.Analog outputs were used to get heart rate and the spo2 exported to outside the unit to a monitoring room.The alarm battery and the battery critical was inactive.Low battery and speaker failures and spo2 pulse were inactive. the unit typically gets turned off and on every day because it only gets used at night.The instructions for swapping out and investigation for wifi issues were sent via email. there was no reported patient outcome.
 
Manufacturer Narrative
Correction: b5, h6 additional information: d9, g3, h3, h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during use, the unit was intermittently malfunctioning with gaps in trend data, there were alarms and it shut itself down.There was error message of system failure and trend data lost.The power and battery were verified, and unit was restarted.Analog outputs were used to get heart rate and spo2 exported to outside the unit to a monitoring room.The alarm battery was critically inactive.It has low battery, speaker failures and spo2 pulse were inactive.The unit typically gets turned off and on during day, because it was only used at night.The monitor settings available were high/low spo2: 100/85 high/low pulse: 170/40 but there were no readings available.There was no patient injury.
 
Manufacturer Narrative
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted there was a poor connection between the two components.The battery failed but was due to normal usage.It was reported that the unit powered down unexpectedly but then turned back on by itself, the product had an issue with the alarm system, the was an issue with the battery, the trend data was lost, the speaker in the monitor was defective, the monitor had a delay in updating the respiratory rates and the monitor was powered on but not providing oximetry readings.The reported issues were confirmed.The most likely cause was traced to a component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NELLCOR
Type of Device
OXIMETER
Manufacturer (Section D)
NELLCOR PURITAN BENNETT IRELAND
micheal collins rd mervue
galway
EI 
Manufacturer (Section G)
NELLCOR PURITAN BENNETT IRELAND
micheal collins rd mervue
galway
EI  
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key16278273
MDR Text Key308618591
Report Number8020893-2023-00083
Device Sequence Number1
Product Code DQA
UDI-Device Identifier10884521182530
UDI-Public10884521182530
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141518
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 07/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPM1000N
Device Catalogue NumberPM1000N
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received07/26/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/13/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age44 YR
Patient SexFemale
-
-