• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION AK 96 MACHINES; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION AK 96 MACHINES; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 110655
Device Problem Inadequate Ultra Filtration (1656)
Patient Problem Abdominal Distention (2601)
Event Date 10/21/2022
Event Type  Injury  
Event Description
It was reported an ak 96 machine experienced a reverse ultrafiltration event.After two hours of treatment the patient "felt oppressed" and gained a little weight.Treatment was immediately stopped, the machine was swapped for a different ak 96 and treatment continued.The symptoms disappeared.The treatment was continued oin a different dialysis machine and the discomfort disappeared.No additional information is available.
 
Manufacturer Narrative
Initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: a service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.The device was evaluated onsite by a local technician.The machine was found working as per specifications.The reported condition was not duplicated,.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AK 96 MACHINES
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - LUND MONITORS
magistratsvagen 16
lund skane lan SE-22 643
SW   SE-22643
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16282845
MDR Text Key309155412
Report Number9616026-2023-00006
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 03/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number110655
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/17/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age64 YR
Patient SexMale
-
-