• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM Back to Search Results
Model Number CAR-505
Device Problems Break (1069); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The cartridge was discarded and not available for evaluation.All devices must meet all quality criteria and manufacturing specifications prior to release.The instructions for use states "make sure mated luer-connectors are secure but do not over-tighten, especially when connections are wet.".
 
Event Description
A report was received on 18 jan 2023 from the nurse at a critical care facility regarding a patient with unspecified pathology, who stated the cartridge luer connector broke during a continuous renal replacement therapy (crrt) treatment and blood could not be returned to the patient on an unspecified date.Additional information was received on 26 jan 2023 from the nurse who stated there was no adverse impact related to the event and no medical intervention required.Although requested, date of event could not be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NXSTAGE SYSTEM ONE
Type of Device
HIGH PERMEABILITY HEMODIALYSIS SYSTEM
Manufacturer (Section D)
NXSTAGE MEDICAL, INC.
350 merrimack street
lawrence MA 01843
Manufacturer (Section G)
MEDIMEXICO S. DE R. L. DE C. V
av. valle imperial no. 10523
parque industrial valle sur
tijuana 22180
MX   22180
Manufacturer Contact
paula rogalski
nxstage medical, inc.
350 merrimack street
lawrence, MA 01843
9784505276
MDR Report Key16282876
MDR Text Key309135114
Report Number3003464075-2023-00012
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133547
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 02/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCAR-505
Device Catalogue NumberCAR-505
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-