• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CDDRA500Q
Device Problems High impedance (1291); Defective Device (2588); Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2023
Event Type  malfunction  
Event Description
It was reported that during an implant procedure, the implantable cardioverter defibrillator (icd) was found to have high pacing impedance and an alleged header anomaly, resulting in inadequate lead insertion into the device header.The icd was not used and another device was used to complete the procedure.
 
Manufacturer Narrative
The reported event of high atrial pacing impedance and displaced atrial contact spring were confirmed.Upon receipt the atrial contact spring was found to have been dislodged and pushed to the tip of the atrial port.The displaced right atrial contact spring is believed to be the cause of the reported event of high atrial impedance measurement.Device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT DR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16305342
MDR Text Key308875557
Report Number2017865-2023-05140
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032003
UDI-Public05415067032003
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDDRA500Q
Device Catalogue NumberCDDRA500Q
Device Lot NumberP000155581
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/29/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/12/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-