• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM Back to Search Results
Model Number CAR-172-C
Device Problem Fluid/Blood Leak (1250)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 01/17/2023
Event Type  malfunction  
Manufacturer Narrative
The cartridge was received for evaluation and revealed a leak in the arterial line.A device history record (dhr) review was conducted for the reported lot number and revealed the product was released for distribution having met quality and manufacturing specifications and requirements.The instructions for use states "check the system for blood and fluid leaks during treatment, and pay close attention to the blood line and access connections".
 
Event Description
A report was received on (b)(6) 2023 from a 44 year old male patient with a medical history including hypertension and end stage renal disease, who stated blood was observed leaking from the cartridge approximately two hours into a hemodialysis treatment on (b)(6) 2023.Additional information was received on 18 jan 2023 from the home therapy nurse (htn) who stated the patient did not experience any symptoms and there was no medical intervention required.Following the event, the patient continues to treat with the nxstage system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NXSTAGE SYSTEM ONE
Type of Device
HIGH PERMEABILITY HEMODIALYSIS SYSTEM
Manufacturer (Section D)
NXSTAGE MEDICAL, INC.
350 merrimack street
lawrence MA 01843
Manufacturer (Section G)
MEDIMEXICO S. DE R. L. DE C. V
av. valle imperial no. 10523
parque industrial valle sur
tijuana 22180
MX   22180
Manufacturer Contact
paula rogalski
nxstage medical, inc.
350 merrimack street
lawrence, MA 01843
9784505276
MDR Report Key16313219
MDR Text Key309369747
Report Number3003464075-2023-00016
Device Sequence Number1
Product Code KDI
UDI-Device IdentifierM535CAR172C0
UDI-Public+M535CAR172C0/$$032421077025/
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140526
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/01/2024
Device Model NumberCAR-172-C
Device Catalogue NumberCAR-172-C
Device Lot Number21077025
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age44 YR
Patient SexMale
Patient Weight147 KG
-
-