MEDTRONIC HEART VALVES DIVISION EVOLUT PRO TRANSCATHETER AORTIC VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
|
Back to Search Results |
|
Model Number EVOLUTPRO-23 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Bacterial Infection (1735); Dyspnea (1816); Endocarditis (1834); Hemorrhage/Bleeding (1888); Rupture (2208); Pseudoaneurysm (2605); Pericardial Effusion (3271)
|
Event Date 01/16/2023 |
Event Type
Injury
|
Event Description
|
Medtronic received information via literature regarding a 69-year-old male patient with coronary artery disease and severe aortic stenosis who underwent transcatheter aortic valve replacement (tavr) using a medtronic evolut pro bioprosthetic valve (unique device identifier numbers not provided).His post-tavr course was complicated with bacteremia, though no infection source was detected, even after a workup including a white blood cell-tagged nuclear scan.Eight months post-tavr, the patient presented to the hospital with altered mental status and shortness-of-breath.A chest computerized tomography (ct) scan revealed an ascending aortic rupture with pseudoaneurysm and a large hemopericardium.Echocardiography showed the hemopericardium as well as trace paravalvular leakage at the bioprosthesis, but no regurgitation.The patient underwent emergent surgery to repair the rupture.During surgery, it was discovered the native aortic root had degenerated from infective endocarditis (ie) involving the bioprosthesis, accounting for the aortic rupture.The decision was made to excise the aortic root and bioprosthetic valve then replace the aortic root utilizing a homograft and reconstruct the coronary arteries.No additional adverse patient effects or product performance issues were reported.
|
|
Manufacturer Narrative
|
Citation: singh s.; et al.Tavr explant with aortic root replacement using a homograft for an aortic rupture.Multimed man cardiothorac surg.(b)(6)2023.Doi: 10.1510/mmcts.2022.101.Pmid: 36644957 earliest date of publication used for date of event.Medtronic products referenced: evolut pro (pma# p130021, product code: npt).Earliest approved product used for product code and pma#.No unique device identifier (serial/lot) numbers were provided; without this information it could not be determined whether these observations have been previously reported.Without return of the product no definitive conclusion can be made regarding the clinical observations.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|