• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INSULET CORPORATION OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM; PUMP, INFUSION, INSULIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSULET CORPORATION OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM; PUMP, INFUSION, INSULIN Back to Search Results
Model Number 18239
Device Problems Crack (1135); Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/06/2023
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned/received to date.If the device is received, a supplemental report will be submitted with the investigation results.We are unable to confirm the reported the reported thermal event.No lot release records were reviewed, as the product lot number was not provided.
 
Event Description
A patient reports their personal diabetes manager (pdm) was hot to the touch while delivering a bolus as well as when it is being charged.In addition the patient reports their pdm has a crack in it from being dropped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM
Type of Device
PUMP, INFUSION, INSULIN
Manufacturer (Section D)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer (Section G)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer Contact
michael spears
100 nagog park
acton, MA 01720
9786007000
MDR Report Key16354509
MDR Text Key309304254
Report Number3004464228-2023-04333
Device Sequence Number1
Product Code LZG
UDI-Device Identifier10385082000009
UDI-Public(01)10385082000009
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K192659
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 02/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number18239
Device Catalogue NumberUSA1-D001-MG-USA1
Was Device Available for Evaluation? No
Date Manufacturer Received02/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age11 YR
Patient SexMale
-
-