• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. ULTRASONIC PROBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. ULTRASONIC PROBE Back to Search Results
Model Number UM-G20-29R-3
Device Problems No Display/Image (1183); Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/15/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The customer reported to olympus, the ultrasonic probe did not produce any ultrasound images.It was also reported that the ultrasonic medium at the tip seemed to leak and adhered to the glove.There did not seem to be any crushing holes in the external appearance of the probe.The issue was found during preparation for use.There were no reports of patient harm.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and correction to b5.Please see updates to b5, h6 and h10.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it¿s likely the ultrasonic medium leaked due to external forces that were applied to the tip sheath which resulted in the tip sheath to puncture.The root cause of this event was unable to be identified.Olympus will continue to monitor field performance for this device.
 
Event Description
This report is being submitted for ultrasonic medium leak.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC PROBE
Type of Device
ULTRASONIC PROBE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16354646
MDR Text Key309497318
Report Number3002808148-2023-01290
Device Sequence Number1
Product Code ITX
UDI-Device Identifier04953170078668
UDI-Public04953170078668
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K001203
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 03/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUM-G20-29R-3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/22/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-