Catalog Number 2C6571 |
Device Problems
Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that the unspecified quantity of continu-flo solution sets were leaking with remicade.The leak was coming from a pin hole on the tubing near the screw end of the catheter.This was observed during patient infusion.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
The events occurred between (b)(6) 2022 to (b)(6) 2023.Device manufacturer address 1: 600 mts.(b)(4) las americas, parque industrial should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
H10: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The devices were discarded; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
|