• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MOMENTUM X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MOMENTUM X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number G138
Device Problems Failure to Convert Rhythm (1540); Inappropriate/Inadequate Shock/Stimulation (1574); Inaccurate Synchronization (1609)
Patient Problem Electric Shock (2554)
Event Date 01/09/2023
Event Type  Injury  
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) delivered inappropriate therapy for atrial fibrillation (af) which was rapidly conducted into the ventricles.On (b)(6) 2023 the patient went into ventricular tachycardia (vt) and therapy was delivered however the shock degraded the rhythm to a fine ventricular fibrillation (vf).The device continued to deliver therapy until exhaustion without conversion.The patient was hospitalized at the time for an unspecified reason and the vf was successfully terminated with an external defibrillator.The patient is currently on extracorporeal membrane oxygenation (ecmo) and possibly has a do not resuscitate order (dnr) in place.The non-boston scientific right ventricular (rv) lead was programmed subthreshold due to high threshold values.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOMENTUM X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16362633
MDR Text Key309387920
Report Number2124215-2023-06194
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589188
UDI-Public00802526589188
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/08/2023
Device Model NumberG138
Device Catalogue NumberG138
Device Lot Number388809
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/19/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/20/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age55 YR
Patient SexMale
-
-