Model Number 50000000E
Device Problem Computer Operating System Problem (2898)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/23/2023
Event Type  malfunction  
Event Description
It was reported that the biomed was getting an alert 61 (non-recoverable system error) on the arctic sun device and would like to send it in for the processor board to be replaced and also would like the 2000 hour preventive maintenance performed while they have it.Per follow up information received via email on 30jan2023, biomed stated that the device was coming in for service and repair.There was no patient involvement.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Manufacturer Narrative
Upon further review, bd has determined that this mdr was reported in error as it was found to be a duplicate of an event previously reported.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that the biomed was getting an alert 61 (non-recoverable system error) on the arctic sun device and would like to send it in for the processor board to be replaced and also would like the 2000 hour preventive maintenance performed while they have it.Per follow up information received via email on 30jan2023, biomed stated that the device was coming in for service and repair.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC SUN® 5000
Type of Device
ARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key16364408
MDR Text Key309465014
Report Number1018233-2023-00829
Device Sequence Number1
Product Code DWJ
UDI-Device Identifier00801741127755
UDI-Public(01)00801741127755
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number50000000E
Device Catalogue Number50000000E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/14/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-