Model Number 795210
Device Problem Unintended System Motion (1430)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/19/2023
Event Type  malfunction  
Manufacturer Narrative
The device was evaluated by an engineer.The engineer adjusted the cable for the affiniti 70 ultrasound system's release handle which resolved the reported issue.The system was returned to service.No further issues were reported.
 
Event Description
A customer reported that their affiniti 70 ultrasound system's articulation arm was not locking in place.The failure occurred outside of clinical use and no patient or user was harmed as a result of the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFINITI 70
Type of Device
SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC
Manufacturer (Section D)
PHILIPS ULTRASOUND
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS ULTRASOUND
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
thuy nguyen
22100 bothell everett highway
bothell, WA 98021
MDR Report Key16365382
MDR Text Key309581007
Report Number3019216-2023-00015
Device Sequence Number1
Product Code IYN
UDI-Device Identifier00884838056251
UDI-Public00884838056251
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132304
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number795210
Device Catalogue Number795210
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-