Model Number 18325
Device Problem Material Split, Cut or Torn (4008)
Patient Problem Hyperglycemia (1905)
Event Date 02/02/2023
Event Type  malfunction  
Event Description
It was reported the patient's blood glucose level rose to 350 mg/dl while wearing the pod.When removed from the infusion site, the pod's cannula was found to be damaged.The patient reported that the infusion site was larger than normal.As treatment the patient replaced the pod.
 
Manufacturer Narrative
The device has not been returned/received to date.If the device is received, a supplemental report will be submitted with the investigation results.It was reported that the cannula had been damaged.This condition could interrupt insulin delivery and contribute to hyperglycemia.Lot release records were reviewed and the product lot met all acceptance criteria.Specifically, a pod is paired to a pdm and put through simulated use testing including communicating with the pdm, deployment, delivering fluid, occlusion detection, and freedom from hazard alarms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM
Type of Device
PUMP, INFUSION, INSULIN
Manufacturer (Section D)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer (Section G)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer Contact
michael spears
100 nagog park
acton, MA 01720
9786007000
MDR Report Key16366316
MDR Text Key309466968
Report Number3004464228-2023-04451
Device Sequence Number1
Product Code LZG
UDI-Device Identifier20385082000051
UDI-Public(01)20385082000051(11)211031(17)230430(10)PD1C10312141
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K192659
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 02/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/30/2023
Device Model Number18325
Device Catalogue NumberBLE-P1-525
Device Lot NumberPD1C10312141
Was Device Available for Evaluation? No
Date Manufacturer Received02/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/31/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age69 YR
Patient SexMale
-
-