Catalog Number 04U42-84 |
Device Problem
Missing Test Results (3267)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/21/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
|
|
Event Description
|
The customer observed an error in open mode for hct, mcv and plt on the cell-dyn ruby analyzer.Qc values were low for all three levels.The error occurred on (b)(6) 2023.Due to the error, results for a patient in intensive care were delayed while the sample was tested on a different analyzer.The customer received a complaint from the physician regarding the delay.The customer did not have any details regarding the patient's health condition.No patient results were available.There was no information available regarding impact to the patient due to the delay.Multiple attempts were made to gather additional information; however, no further details are available at this time.There is no known patient harm due to the delay of results.
|
|
Manufacturer Narrative
|
Data was not provided by the customer for review.Field service performed an optics bench alignment procedure which resolved the complaint issue.Trending review determined no trend for the issue for the product.Device history record review did not show any potential non-conformances, or deviations.Labeling was reviewed and found to adequately address the issue under review.Based on the investigation, no deficiency was identified.
|
|