MAUDE Adverse Event Report: MASIMO - 15750 ALTON PKWY RAD-G PULSE OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 15750 ALTON PKWY RAD-G PULSE OXIMETER Back to Search Results
Model Number 27977
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/20/2023
Event Type  malfunction  
Manufacturer Narrative
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.Zip code is as follows: (b)(6).
 
Event Description
The customer reported the device will turn off on it's own.No consequences or impact to patient were reported.
 
Manufacturer Narrative
Additional manufacuring narrative: the returned device was evaluated.During evaluation the device passed all visual and functional testing.The unit was found to visually and audibly alarm during alarm conditions.The unit was determined to be functioning as designed.
 
Event Description
The customer reported the device will turn off on its own.No consequences or impact to patient were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAD-G PULSE OXIMETER
Type of Device
OXIMETER
Manufacturer (Section D)
MASIMO - 15750 ALTON PKWY
15750 alton pkwy
irvine CA 92618
Manufacturer (Section G)
MASIMO - MEXICALI
industrial vallera de mexicali calzada del oro, no.2001
mexicali, baja california 21600
MX   21600
Manufacturer Contact
matofa fautua
15750 alton pkwy
irvine, CA 92618
9498223661
MDR Report Key16374613
MDR Text Key309527467
Report Number3019388613-2023-00027
Device Sequence Number1
Product Code DQA
UDI-Device Identifier00843997000666
UDI-Public00843997000666
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K201770
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative,Distributor
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/14/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number27977
Device Catalogue Number9849
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/09/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-