EDWARDS LIFESCIENCES EDWARDS SAPIEN 3 TRANSCATHETER; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 9750TFX26
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Rupture (2208)
Event Date 01/24/2023
Event Type  Death  
Manufacturer Narrative
Per the instructions for use (ifu), cardiovascular injuries such as perforation or dissection of vessels, ventricle, myocardial or valvular structures, and hematoma are known potential adverse events associated with the overall thv procedure and may require intervention.According to the thv training manuals, risk factors for aortic dissection, hematoma, or annular rupture during the thv procedure include significant thv over sizing, severely obliterated sinuses of valsalva, porcelain aorta and/or presence of bulky calcification, and narrow calcified stj.In addition, advanced age, female gender, small body weight, and steroid dependency can also be contributing factors.The sapien valve relies on native valve calcium to securely anchor to the annulus.Despite this beneficial aspect of calcium, bulky calcium can increase the risk of calcific nodule displacement into the vasculature, which can lead to vascular injury.At times, the extent and distribution of calcium can impair ease of delivery of the valve, correct positioning of the valve, deployment of the valve, and procedural success.In this case, there was no allegation or indication a product malfunction contributed to this adverse event.Investigation results suggest both patient factors and procedural factors may have caused or contributed to this event.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required.Device remains implanted.
 
Event Description
As reported, it was a case of an implant of a 26mm sapien 3 ultra thv in aortic position by transfemoral approach.Due to a high calcium score of the native valve (+7000) and critical aortic stenosis, the native valve was pre-dilated.While the deployment balloon was fully inflated, the operator performed a power injection which resulted in a cardiac injury with subsequent pericardial effusion and tamponade.Patient was transferred to open chest surgery.Analysis of the pre-dilation fluoroscopy showed that at that point a dissection/annular rupture happened and was exacerbated later with the power injection of the balloon of the delivery system.Patient passed away.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS SAPIEN 3 TRANSCATHETER
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key16382186
MDR Text Key309612664
Report Number2015691-2023-10827
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P140031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number9750TFX26
Device Catalogue Number9750TFX26A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/24/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death; Life Threatening; Required Intervention;
Patient Age82 YR
Patient SexFemale
-
-