Catalog Number 10802688
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/23/2023
Event Type  malfunction  
Manufacturer Narrative
Medical device expiration date: unknown.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event Description
It was reported that the bd alaris¿ smartsite¿ gravity set tubing broke off below the drip chamber while being primed.The following information was provided by the initial reporter: "bd smartsite gravity set broke off when connected to iv fluid and primed.Tubing broke off just below the drip chamber.".
 
Event Description
It was reported that the bd alaris¿ smartsite¿ gravity set tubing broke off below the drip chamber while being primed.The following information was provided by the initial reporter: "bd smartsite gravity set broke off when connected to iv fluid and primed.Tubing broke off just below the drip chamber.".
 
Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 09-feb-2023.H6: investigation summary one sample was received and tested by our quality team.The separation was immediately noticed and the complaint of drip chamber separation was verified.The tubing was analyzed under microscope and there was a clear lack of solvent where it was to be inserted into the drip chamber.The root cause is due to an improper use of assembly equipment to apply solvent to tubing.A device history record review could not be performed because a lot number was not provided by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS¿ SMARTSITE¿ GRAVITY SET
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key16382747
MDR Text Key309640456
Report Number9616066-2023-00213
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10885403234439
UDI-Public10885403234439
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K820278
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number10802688
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/27/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-