ARROW INTERNATIONAL LLC ARROW PI PICC: 1-L 4FR X 40CM 80CM HNIT WIRE; CATHETER,INTRAVASCULAR,THERAPE
|
Back to Search Results |
|
Catalog Number PR-34041-NM |
Device Problem
Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/17/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported "after placement of the catheter, the piece of the cut catheter was found next to the patient bed.Therefore, the catheter was replaced with a new one.No harm to the patient was reported." the patient's condition was reported to be fine.
|
|
Manufacturer Narrative
|
(b)(4).The customer provided one photo for evaluation.The image shows a picc catheter that appears stretched and torn.The customer also returned one s-l picc for analysis.Visual inspection revealed that the catheter was fully cut into two pieces at the level of the clamp catheter/clamp fastener.Both separation edges are rough and jagged.The catheter appeared stretched and deformed at the points of separation, and the markings near the separation points are stretched/deformed.The appearance of the damage indicates that the catheter was subjected to undue force.The catheter pieces measured 13 3/4" and 4 1/2", totaling 18 1/4", (b)(4).The markings on both segments of the catheter appeared to line up according to specification.The shorter piece of the catheter was observed to be significantly stretched, which corresponds to the out of spec measurement.Functional inspection of the catheter could not be adequately performed due to the damage.A manual tug test confirmed the extension line was fully secured to the luer hub.The customer did not provide a lot number; therefore, a device history record review was performed based upon a lot number from the sales history data of the customer.No relevant findings were identified.The instructions for use (ifu) provided with this kit warns the user, "do not apply excessive force in placing or removing catheter or guidewire.Excessive force can cause component damage or breakage.If damage is suspected or withdrawal cannot be easily accomplished, radiographic visualization should be obtained and further consultation requested.Minimize catheter manipulation throughout procedure to maintain proper catheter tip position." the customer report of a cut/torn catheter was confirmed through complaint investigation of the returned sample.Visual inspection revealed the catheter was fully cut into two segments.The separation edges appeared rough and jagged, and the catheter appeared stretched at the point of separation.The catheter met all relevant dimensional requirements, and a device history record review based on sales history did not reveal any evidence to suggest a manufacturing related issue.Based on the customer report that the catheter was damaged during use and the appearance of the damage, unintentional use error likely caused or contributed to this event.Teleflex will continue to monitor and trend for complaints of this nature.
|
|
Event Description
|
It was reported "after placement of the catheter, the piece of the cut catheter was found next to the patient bed.Therefore, the catheter was replaced with a new one.No harm to the patient was reported." the patient's condition was reported to be fine.
|
|
Search Alerts/Recalls
|
|
|