Model Number WNDARM
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 01/18/2023
Event Type  Injury  
Manufacturer Narrative
Based on the information provided, it cannot be determined that the alleged bleeding event requiring suturing is related to the activ.A.C.¿ ion progress¿ remote therapy monitoring system.The patient is on anticoagulation therapy and has a history of bleeding that required suture ligation prior to v.A.C.® therapy placement.Multiple unsuccessful attempts have been made to obtain additional clinical information.The device passed quality control checks before patient placement, and a device evaluation after placement could not be performed.Device labeling, available in print and online, states: contraindications do not place foam dressings of the v.A.C.® therapy system directly in contact with exposed blood vessels, anastomotic sites, organs, or nerves.Warnings bleeding: with or without using v.A.C.® therapy, certain patients are at high risk of bleeding complications.The following types of patients are at increased risk of bleeding, which, if uncontrolled, could be potentially fatal: patients who have weakened or friable blood vessels or organs in or around the wound as a result of, but not limited to: suturing of the blood vessel (native anastomosis or grafts) / organ, infection, trauma, radiation, patients without adequate wound hemostasis, patients who have been administered anticoagulants or platelet aggregation inhibitors,and patients who do not have adequate tissue coverage over vascular structures.If v.A.C.® therapy is prescribed for patients who have an increased risk of bleeding complications, they should be treated and monitored in a care setting deemed appropriate by the treating physician.If active bleeding develops suddenly or in large amounts during v.A.C.® therapy, or if frank (bright red) blood is seen in the tubing or in the canister, immediately stop v.A.C.® therapy, leave dressing in place, take measures to stop the bleeding and seek immediate medical assistance.The v.A.C.® therapy units and dressings should not be used to prevent, minimize or stop vascular bleeding.Protect vessels and organs: all exposed or superficial vessels and organs in or around the wound must be completely covered and protected prior to the administration of v.A.C.® therapy.Always ensure that v.A.C.® foam dressings do not come in contact with vessels or organs.Use a thick layer of natural tissue should provide the most effective protection.If a thick layer of natural tissue is not available or is not surgically possible, multiple layers of non-adherent dressing material may be considered as an alternative, if deemed by the treating physician to provide a complete protective barrier.If using non-adherent materials, ensure they are secured in a manner that will maintain their protective position throughout therapy.Disclaimer: this information is submitted pursuant to 21 cfr 803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a kinetic concepts, inc.Product malfunctioned, is defective or has caused serious injury.
 
Event Description
On (b)(6) 2023, the following information was provided to kci by the patient: the physician lowered the activ.A.C.¿ ion progress¿ remote therapy monitoring system pressure setting to 100mmhg allegedly due to bleeding.The physician applied sutures and no further bleeding occurred.No additional information was provided.On (b)(6) 2022, the device was tested per quality control procedure by the kci service center, and the unit passed the quality control checks and met specifications.On (b)(6) 2023, the device was placed with the patient.Multiple unsuccessful attempts have been made to retrieve the device, therefore, a device evaluation could not be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIV.A.C.¿ ION PROGRESS¿ REMOTE THERAPY MONITORING SYSTEM
Type of Device
OMP
Manufacturer (Section D)
KINETIC CONCEPTS, INC.
san antonio TX 78249
Manufacturer Contact
steven jackson
6203 farinon drive
san antonio, TX 78249
2102556438
MDR Report Key16390708
MDR Text Key309729908
Report Number3009897021-2023-00011
Device Sequence Number1
Product Code OMP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201571
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/16/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberWNDARM
Device Catalogue Number420096
Was Device Available for Evaluation? No
Date Manufacturer Received01/18/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age61 YR
Patient SexMale
-
-