MAUDE Adverse Event Report: MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN® 5000; ARCTIC SUN DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN® 5000; ARCTIC SUN DEVICE Back to Search Results
Model Number 50000000E
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/31/2023
Event Type  malfunction  
Event Description
It was reported that the arctic sun device booted to an alarm 74 (non-recoverable system error).Biomed explained this was indicative of a t2 (outlet control temperature) thermistor failure.Per follow up information received via email on (b)(6) 2022, there was no patient involvement.Per sample evaluation results received on (b)(6) 2023, the root cause was isolated to a faulty t2 (outlet control temperature) thermistor on the tank harness.It was reported the hot side connection between the power inlet module and the ac main voltage circuit card showed signs of electrical overstress.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Manufacturer Narrative
The reported issue was confirmed.The device was evaluated and the reported issue was confirmed as it was noted the hot side connection between the power inlet module and the ac main voltage circuit card shows signs of electrical overstress.Preventatively replaced ac main voltage circuit card.Completed the 2000-hour pm procedure on the device.Serviced the circulation pump and mixing pump, replaced the heater lot 1902 with lot 2205, and replaced both manifold o-rings and drain valves the device was put through a functional check.The device was heated to 42°c and cooled to 4°c and was stable at 28°c with a flow of 1.8 l/m with -7.0 psi in bypass mode.The device was placed on acats (automated calibration and test system).An electrical safety test was performed.A 45 hour burn in was performed.A final inspection was performed.The arctic sun 5000 passed all performance testing, calibration and electrical safety tests and is functioning properly and ready for use.The overheating of the wires was assessed by the investigation tasks and used to complete the root cause evaluation using the fishbone methodology (refer to ac cca power inlet module powerpoint for fishbone diagram).It was concluded that the following was the root cause: supplier ¿ root cause : inadequate verification and validation activities of the crimping process, single pull test did not provide stability of process, evidence was not provided when requested for maintenance of records or crimp tools, no crimp cross-sections provided.The dhr review is not required and the labeling review is not required because labeling could not have prevented this issue.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that the arctic sun device booted to an alarm 74 (non-recoverable system error).Biomed explained this was indicative of a t2 (outlet control temperature) thermistor failure.Per follow up information received via email on 02-aug-2022, there was no patient involvement.Per sample evaluation results received on 31-jan-2023, the root cause was isolated to a faulty t2 (outlet control temperature) thermistor on the tank harness.It was reported the hot side connection between the power inlet module and the ac main voltage circuit card showed signs of electrical overstress.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC SUN® 5000
Type of Device
ARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key16410508
MDR Text Key310134510
Report Number1018233-2023-01048
Device Sequence Number1
Product Code DWJ
UDI-Device Identifier00801741127755
UDI-Public(01)00801741127755
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number50000000E
Device Catalogue Number50000000E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-