• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number MCP00706291#ROTAFLOW PUMP MODULE
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2023
Event Type  malfunction  
Event Description
The event occurred in china.It was reported that the error message ¿head error¿ occurred during start-up of the rotaflow console.No patient was involved.After testing is was found a problem with the control board and drive.No harm to any person has been reported.Complaint id: (b)(4).
 
Manufacturer Narrative
The investigation is ongoing.Further information has been requested but has not yet been received.A follow up medwatch will be submitted when additional information becomes available.
 
Manufacturer Narrative
The event occurred in china.It was reported that the error message ¿head error¿ occurred during start-up of the rotaflow console.No patient was involved.No harm to any person has been reported.A getinge service technician investigated the affected rotaflow console (rfc) with s/n (b)(6) and the rotaflow drive (rfd) with s/n (b)(6).The technician was able to confirm the reported "head error".The rfc control board pcba (printed circuit board assembly) kit (article number 701034051) on the rfc will be replaced.The rotaflow drive (rfd) with s/n (b)(6) was sent back to manufacturer for repair.During the investigation by the getinge service department on 2023-06-14 the reported "head error" could be reproduced.Thus, the drive was sent to the supplier emtec for repair.On 2023-07-05 the supplier emtec was able to reproduce the reported failure "head error" and the root cause could be determined as misuse of the device, which lead to a damage of the electronic parts.These parts were replaced by the supplier.After functional test at the getinge service department on 2023-07-14 the device was sent back to the user.The following most possible root cause could be determined for the head error: the head error is caused by the hot plug.When the device is in operation and the power plug is plugged in or out the head error occurs and the rota flow drive and / or the control board is damaged.As a result the rota flow drive and / or the control board has to be replaced.Based on these investigation results the reported failure "head error" could be confirmed.Rotaflow console: the review of the non-conformities was performed on 2023-02-22 and during the period of 2019-09-12 to 2023-02-22 does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences are unlikely.The rotaflow console in question was produced in 2019-09-12.Rotaflow drive: the review of the non-conformities was performed on 2023-10-20 and during the period of 2019-09-23 to 2023-10-20 does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences are unlikely.Rotaflow drive: the rotaflow drive in question was produced in 2019-09-23.In order to avoid reoccurrence of the reported failure, the customer will be informed by the getinge sales and service unit (ssu) to follow the chapter in the instruction for use heart-lung support system rotaflow system| 4.4 | en | 15.Chapter 4.1.3: switch off the rotaflow console on/off switch before connecting the rotaflow drive to or disconnecting it from the rotaflow console.Otherwise the rotaflow console may be damaged.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
Complaint id:(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key16425374
MDR Text Key310337926
Report Number8010762-2023-00097
Device Sequence Number1
Product Code KFM
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K991864
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMCP00706291#ROTAFLOW PUMP MODULE
Device Catalogue Number701046405
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/12/2019
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-