MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS SAPIEN 3 ULTRA TRANSCATHETER HEART VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS SAPIEN 3 ULTRA TRANSCATHETER HEART VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 9750TFX26
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Stroke/CVA (1770)
Event Date 02/01/2023
Event Type  Death  
Manufacturer Narrative
Per the instructions for use (ifu), permanent or transient neurological events such as transient ischemic attack (tia) and stroke are known potential adverse events associated with the thv procedure and the use of the edwards thv devices.According to the literature review, and as documented in a clinical technical summary written by ew, stroke is recognized in the literature as a well-known complication in a small number of patients undergoing thv.Risk factors correlating with several patient co-morbidities have been identified.Although in many cases the root cause of the event is unable to be determined, strokes during thv are undoubtedly multifactorial, the dominant etiology likely being intra-procedure embolic events.A transcranial doppler study during thv demonstrated that most procedural embolic events occurred during balloon valvuloplasty, manipulation of catheters across the aortic valve, and valve implantation.An analysis in patients undergoing valve surgery revealed four baseline characteristics and two procedural events that were associated with early post-procedure stroke: female sex, ef < 30%, diabetes, age older than 70 years, bypass procedure time> 120 min, and calcification of the ascending aorta.Predictors of late stroke have included female sex, age older than 75 years, atrial fibrillation, and a history of or current smoking.There were no significant differences in the frequency of late strokes between thv and avr patients.After thv, there appears to be a more considerable proportion of early strokes occurring < 24 h post-procedure, but thv patients with multiple co-morbidities are probably at higher risk of both early and late strokes.In this case, there was no allegation or indication a product malfunction contributed to this adverse event.Investigation results suggest/indicate calcification came loose and ended up in the patient's brain caused or contributed to this event.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required.The device remains implanted.
 
Event Description
As reported, it was a case of a 26mm sapien 3 ultra valve, aortic position by transfemoral approach.After a successful valve implantation, the patient experienced dropout symptoms on the right side and was confused.A ct scan was performed, and it was noted that a chunk of the calcification came loose and ended up in the patient's brain.An attempt to remove the calcification was performed but was unsuccessful.The patient expired.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS SAPIEN 3 ULTRA TRANSCATHETER HEART VALVE
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key16427669
MDR Text Key310068763
Report Number2015691-2023-11008
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
P140031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number9750TFX26
Device Catalogue Number9750TFX26A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/01/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient SexMale
-
-