• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION EUROPEAN VALVE SET, EM2400; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION EUROPEAN VALVE SET, EM2400; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938724E
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/27/2023
Event Type  malfunction  
Manufacturer Narrative
Initial reporter address: thorncliffe park brookdale road should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a em2400 valve set leaked.During compounding, port 19 (ascorbic acid) was pulling into port 22 (cernevit).There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to d9, h3, h4, h6 and h10.H4: the lot was manufactured from july 18, 2022 to july 19, 2022.H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.System level testing was performed and the reported issue was not observed.The valve set was analyzed at various points throughout the prime and verify process and pump calibration process.A leak was not verified on all inlet ports during ui flush after prime and pump calibration delivery process.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EUROPEAN VALVE SET, EM2400
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt 001 mz 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16431011
MDR Text Key310339265
Report Number1416980-2023-00645
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412475776
UDI-Public(01)00085412475776
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberH938724E
Device Lot Number60381983
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/23/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-