Event Description
|
Device report from hong kong reports an event as follows: it was reported that on (b)(6) 2023, the blade that attached to the impactor could not go through the outer sleeve well.There were no reported patient consequences.This report is for a 2.0mm ti matrixmandible screw self-tapping 14mm.This is report 1 of 1 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device available for evaluation: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The initial reporter is a synthes employee.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
There was no surgical delay.Another piece of blade and instrument was used immediately.The surgery was successfully completed.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h6: device problem code added.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D10: concomitant device added.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: a manufacturing record evaluation was performed for the finished device lot and no non-conformance was identified.Product code: 04.027.050s.Lot number: 36p3048.Manufacturing site: jabil bettlach.Release to warehouse date: 21/01/2020.Expiry date: 01/01/2030.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device and from provided photographs.Visual analysis of the returned sample revealed that pfna-ii blade l75 tan was chipped from the tip and sleeve, fragments were not returned.Additionally, the screw was worn from the recess, this condition was most likely created while tightening the locking mechanism.As per visual inspection of the picture, the blade was observed to be assembled with the pfna blade impactor and halfway through the protection sleeve.A dimensional inspection was performed for the pfna-ii blade l75 tan and met specifications.A functional test was performed, a screwdriver was used to adjust the locking screw, sleeve was able to move freely even when the screw was tightened.The complaint condition was able to be replicated.An interaction test was unable to be performed as the impactor and protection sleeve visible in the photo provided were not returned, hence this condition cannot be confirmed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the pfna-ii blade l75 tan would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.H4: device manufacture date corrected.
|
|