• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INOGEN CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INOGEN CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR Back to Search Results
Model Number G141
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Inappropriate/Inadequate Shock/Stimulation (1574); Battery Problem (2885)
Patient Problem Electric Shock (2554)
Event Date 02/05/2023
Event Type  malfunction  
Event Description
It was reported that cardiac resynchronization therapy defibrillator (crt-d) had triggered code 1007 due to the capacitor charge time exceeding limitations along with the battery capacity being depleted.A code 1004 was suspected, which was indicative of a short circuit condition.Additionally, the patient indicated that received some electric shocks.Troubleshooting options were discussed and recommended.At this time, the crt-d remains in service and no adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16436215
MDR Text Key310162884
Report Number2124215-2023-08677
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/25/2017
Device Model NumberG141
Device Catalogue NumberG141
Device Lot Number110378
Was Device Available for Evaluation? No
Date Manufacturer Received02/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/25/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexFemale
-
-