This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms#: (b)(4).No causes or potential causes of the customers reported problem were found during the review of service and repair records.A product sample was received for evaluation.Visual and functional testing were performed.Visual inspection showed the top case was badly chipped on corners and cracked by tubing guides, the bottom case was cracked by l-bracket, the right plunger case was chipped, outside of pump covered in fluid contamination, and there was some fluid ingression present.A review of the event history log identified no alarms related to the reported problem.During functional testing was unable to duplicate the power up problem but did find fluid ingression on interconnect board and main board.Visual inspection confirmed the top case was physically damaged.The root cause of the issue was customer induced via fluid ingression into the main body of the pump as a result of either the customer's improper cleaning of the pump, or the customer's introduction of excessive fluids to the pumps surface.The root cause of physical damage was due to customer impact.Replaced interconnect board and main board.Replaced top case.Performed power up process, preventative maintenance and all functional tests which passed.
|