Device Problems
Device Alarm System (1012); Air/Gas in Device (4062)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/01/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the pump alarmed 'no disposable- clamp tubing.'upon restart, pump alarmed no disposable pump will not run.' settings reviewed, pump turned off, and tubing clamped.Cassette removed and tubing massaged while pressing green flow stop down.Air bubbles were present in the cassette tubing.Cassette was tapped on a hard surface and attached.Pump turned on, settings reviewed and upon restart, alarm returns.Process repeated and alarm persisted.Patient not affected.No patient injury reported.
|
|
Manufacturer Narrative
|
Device identification (udi) is unknown; no further information was provided.No lot number was provided; therefore, a device history record (dhr) review could not be conducted.No product was returned; therefore, no visual and functional tests were performed, the reported complaint could not be confirmed, and the root cause could not be determined.
|
|