MEDTRONIC HEART VALVES DIVISION EVOLUT FX VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
|
Back to Search Results |
|
Model Number EVOLUTFX-26 |
Device Problem
Device Dislodged or Dislocated (2923)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/22/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that during implant of this transcatheter bioprosthetic valve, the valve dislodged and an additional medtronic transcatheter valve was implanted.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Additional information was received that a pre-balloon aortic valvuloplasty (bav) was performed.It was reported that the nose cone possibly pulled on the top of the valve frame causing it to dislodge aortic.Prior to the dislodgement, the implant depth was 2 millimeter (mm) on the non-coronary cusp (ncc) and 2 mm on the left coronary cusp (lcc).Following the dislodgement, the implant depth was 15 mm on the ncc and 15 mm on the lcc.Section d information references the main component of the system.Other relevant device(s) are: product id d-evolutfx-2329 lot 0011541578 use by date 2024-12-06 udi (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Conclusion: dislodgement of the valve by the distal tip is related to operator technique or experience.The device instructions for use instructs ¿under fluoroscopic guidance, confirm that the catheter tip is coaxial with the inflow portion of the bioprosthesis¿ prior to drawing the delivery catheter system (dcs) tip through the valve.With the limited information available and without procedural images for review, a conclusive root cause of the dislodgement event cannot be confirmed.Events of this type do not indicate a device malfunction or a failure to meet manufacturing specifications.Updated: h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|