Model Number CDHFA500Q
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/02/2023
Event Type  malfunction  
Event Description
During an implant procedure when attempting to connect the left ventricular (lv) lead it was noted the set screw was stripped in the device header.A new device was used to resolve the event.The patient was stable.
 
Manufacturer Narrative
The reported event of set screw anomaly was confirmed.The device was above the elective replacement indicator (eri) upon receipt.Analysis revealed the left ventricular set screw was partially stripped and contained septum material inside the hex cavity.This material in the hex cavity prevented full insertion of the torque driver and was the cause of the reported event.The set screw anomaly was consistent with having occurred during the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16449501
MDR Text Key310332143
Report Number2017865-2023-10987
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032010
UDI-Public05415067032010
Combination Product (y/n)N
Reporter Country CodeMN
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2023
Device Model NumberCDHFA500Q
Device Catalogue NumberCDHFA500Q
Device Lot NumberP000130759
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/11/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-