MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT HF; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT HF; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CDHFA500Q
Device Problem Over-Sensing (1438)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/29/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented remotely via merlin.Net.Upon review, the implantable cardioverter defibrillator exhibited post paced t-wave over-sensing.No intervention was performed.No patient consequences.
 
Event Description
New information notes that the implantable cardioverter defibrillator exhibited post paced t wave over-sensing episodes again.No intervention was performed.No patient consequences.
 
Event Description
New information received notes that programming changes were made prior to the most recent post-paced t-wave over-sensing event to resolve the last event.Programming changes were made again to resolve the most recent event.There were no patient consequences.
 
Event Description
New information notes that the implantable cardioverter defibrillator exhibited post paced t wave over-sensing episodes again.No intervention was performed.No patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16453387
MDR Text Key310354067
Report Number2017865-2023-11322
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032010
UDI-Public05415067032010
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup,Followup
Report Date 05/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2023
Device Model NumberCDHFA500Q
Device Catalogue NumberCDHFA500Q
Device Lot NumberS000084721
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/24/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/15/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
LEFT VENTRICLE LEAD; RIGHT ATRIAL LEAD; RIGHT VENTRICLE LEAD
Patient Age73 YR
Patient SexMale
Patient Weight77 KG
-
-