Model Number 9750TFX23
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Obstruction/Occlusion (2422)
Event Date 09/27/2022
Event Type  Injury  
Event Description
As reported by a spain affiliate, it was a case of a 23mm sapien 3 ultra valve in the aortic position by transfemoral approach.There was low origin of the coronary arteries with protrusion of the native aortic leaflets towards the ostium of the left common trunk, which required the implantation of a stent with chimney technique.Event was resolved.
 
Manufacturer Narrative
Per the instructions for use (ifu), coronary flow obstruction is a known potential adverse event associated with the tavr procedure.The ifu cautions that safety and effectiveness have not been established for patients with bulky calcified aortic valve leaflets in close proximity to coronary ostia.In addition, it warns that caution should be exercised in implanting a bioprosthesis in patients with clinically significant coronary artery disease.Coronary obstruction can result in myocardial ischemia or infarction due to obstruction of the coronary blood flow and may require intervention (e.G.Percutaneous coronary intervention (pci)).Multiple patient factors could put the patient at risk for coronary flow obstruction during the thv procedure, including significant underlying coronary artery disease and bulky calcification of the native leaflets and root.Displacement of calcium deposits with embolization of debris into one of the arteries, or aortic dissection with continuity of the rupture into the intima of one of the coronary ostia, can result in this complication.Additionally, minimal distance between the native annulus and the coronary ostia, bulky calcification, long native leaflets, and obliterated coronary sinuses.Procedural factors such as torn native leaflet during bav, plaque shift, deployment of the bioprosthetic heart valve too aortic, and significant valve over sizing could also contribute to this complication.The edwards thv training manuals advise the operator on pre-procedure assessment of the aortic valve, root, and coronary anatomy.Physicians are extensively trained by edwards before they are qualified to use the transcatheter heart valve (thv).Training includes patient screening, device preparation, approach, deployment, imaging, procedure-specific training manuals, and proctored procedures.The physician is instructed to evaluate this risk early in the patient screening process in all patients the following factors should be considered: degree of calcification on leaflets, annulus to coronary ostia distance, length of the valve leaflet, width of the valsalva sinuses, movement of the leaflets during patency of coronaries during bav, and expanded height of the intended thv.The training manuals also provide the following tips for detecting risk for left main occlusion: (1) aortogram or tee before thv implantation to reveal bulky calcified leaflets; (2) during pre-dilatation, note bulky calcification on valve moving towards ostium on left main; and (3) consider aortogram during valvuloplasty to assess coronary flow.The edwards thv manuals also advise the operator on pre-procedure assessment of the aortic valve, root, and coronary anatomy.Physicians are extensively trained by edwards before they are qualified to use the transcatheter heart valve (thv).Training includes patient screening, device preparation, approach, deployment, imaging, procedure-specific training manuals, and proctored procedures.In this case, there was no allegation or indication a product malfunction contributed to this adverse event.Investigation results suggest that patient factors (low origin of the coronary arteries with protrusion of the native aortic leaflets towards the ostium of the left common trunk) may have contributed to the lmt obstruction.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required.Device remains implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS SAPIEN 3 ULTRA
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key16454663
MDR Text Key310371320
Report Number2015691-2023-11155
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
P140031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Study,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/15/2024
Device Model Number9750TFX23
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/02/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/16/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age71 YR
Patient SexFemale
Patient Weight51 KG
-
-