Model Number 0500318E
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/17/2023
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event Description
A user facility clinic manager reported an internal dialyzer blood leak that occurred five minutes into a patient¿s hemodialysis (hd) treatment.The blood leak was visually observed in the dialysate compartment of the dialyzer.The machine, a fresenius 2008t machine, alarmed appropriately with a blood leak alarm.A blood leak test strip was not used.No visible damage was observed on the dialyzer.After the machine alarmed, the treatment was paused.The patient¿s blood was not returned; estimated blood loss (ebl) was 200 ml.Upon followup the registered nurse (rn) confirmed there was no patient injury, no adverse effects were experienced, and no medical intervention was required as a result of the reported event.The patient completed treatment after being resetup with new supplies on a different machine.The complaint device was reportedly not available to be returned for manufacturer evaluation.
 
Manufacturer Narrative
Plant investigation: the reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation.During the lot history review it was noted that there were no other complaints of any kind reported against the lot.A production records review was performed on the reported lot.The production record review showed there was one non-conformance (nc) in the production of this lot.It is unrelated to the reported complaint event.There was no indication of product nonacceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event.The lot met all release criteria.A definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device.Therefore, the complaint is not confirmed.Continuous improvement is of the utmost importance to fresenius medical care as we strive to provide dialysis products of the highest quality to our patients.Reports of leaking product are investigated both individually as complaints as well as via the non-conformance/capa program, in order to assess and improve our products and processes.Capas for vision systems and blood leak reduction are recent examples of leak related investigations directed at an overall reduction in dialyzer leaks.
 
Event Description
A user facility clinic manager reported an internal dialyzer blood leak that occurred five minutes into a patient¿s hemodialysis (hd) treatment.The blood leak was visually observed in the dialysate compartment of the dialyzer.The machine, a fresenius 2008t machine, alarmed appropriately with a blood leak alarm.A blood leak test strip was not used.No visible damage was observed on the dialyzer.After the machine alarmed, the treatment was paused.The patient¿s blood was not returned; estimated blood loss (ebl) was 200 ml.Upon followup the registered nurse (rn) confirmed there was no patient injury, no adverse effects were experienced, and no medical intervention was required as a result of the reported event.The patient completed treatment after being resetup with new supplies on a different machine.The complaint device was reportedly not available to be returned for manufacturer evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLUX 180NRE DIALYZER FINISHED ASSY.
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer (Section G)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key16457224
MDR Text Key310427889
Report Number0001713747-2023-00151
Device Sequence Number1
Product Code KDI
UDI-Device Identifier00840861100156
UDI-Public00840861100156
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162488
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0500318E
Device Catalogue Number0500318E
Device Lot Number22PU02019
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Date Manufacturer Received03/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/27/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
FRESENIUS 2008T MACHINE; FRESENIUS 2008T MACHINE; FRESENIUS BLOODLINES; FRESENIUS BLOODLINES
Patient Age66 YR
Patient SexFemale
Patient Weight88 KG
-
-