• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET LEAD DOUBLE BEND, 86 CM; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET LEAD DOUBLE BEND, 86 CM; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1457Q/86
Device Problems Break (1069); Failure to Capture (1081); High impedance (1291); Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/10/2023
Event Type  malfunction  
Event Description
It was reported that during implant.Device interrogation revealed high pacing impedance, failure to sense, failure to capture due to insulation breach on the left ventricular (lv) lead.The physician elected to explant and replace the lv lead.The patient was in stable condition.
 
Manufacturer Narrative
The reported events were high pacing lead impedance, insulation damage, failure to sense and failure to capture.As received, a complete lead was returned in one piece.The reported event of insulation damage was confirmed.The cause of the reported event was found to the lead insulation damage, consistent with procedural damage.No other anomalies were seen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET LEAD DOUBLE BEND, 86 CM
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16464828
MDR Text Key310489478
Report Number2017865-2023-11514
Device Sequence Number1
Product Code OJX
UDI-Device Identifier05414734510189
UDI-Public05414734510189
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1457Q/86
Device Catalogue Number1457Q/86
Device Lot NumberA000130508
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
2088TC/52 ATRIAL; CDHFA500Q; LDA210Q/58 RIGHT VENTRICLE
Patient Age65 YR
Patient SexMale
-
-