The received x-ray was reviewed and the reported necrosis can be confirmed.Only based on this picture it is not possible to determine if the necrosis is in relation with the implants, especially as the devices were implanted for more than 3 years.A device inspection was not possible since the affected device was not returned, therefore no further evaluation is possible.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|