Model Number UHI-4
Device Problems Erratic or Intermittent Display (1182); Pressure Problem (3012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, the user¿s report was not confirmed.The unit was checked and there were no visual nor mechanical abnormalities noted with the device.If additional information becomes available following the device evaluation, a supplemental report will be filed.
 
Event Description
The customer reported that the led on the front panel display of his olympus high flow insufflation unit, was not lighting up.According to the initial reporter, the display is dark and intermittent.Reportedly, the intended procedure was completed.Though requested, specific details were not provided.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the approved final investigation.Section b5 was corrected with information that was inadvertently left out of the initial report.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and since no device malfunction was confirmed during evaluation, the definitive root cause of the display and insufflation issue could not be determined.Olympus will continue to monitor field performance for this device.
 
Event Description
The event occurred during a therapeutic procedure.Two towers were used during procedure.The other tower was completed by switching to the high flow insufflation unit.The device was inspected before use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16529310
MDR Text Key311575941
Report Number3002808148-2023-02419
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2023
Was the Report Sent to FDA? No
Date Manufacturer Received04/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/23/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-