MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MICRORAPTOR KNOTLESS SA PEEK; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MICRORAPTOR KNOTLESS SA PEEK; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72205021
Medical Device Problem Codes Break (1069); Fracture (1260); Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the execution of an arthroscopy procedure, when opening the microraptor knotless peek implant, the proximal part of the anchor was missing, then, it was noticed that the anchor was badly assembled.In addition, the anchor in its distal part was somewhat fractured.The procedure was successfully completed with a non-significant surgical delay using a back-up device.No further complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: case-(b)(4).
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).H10.H3, h6: the reported device was received for evaluation.A visual inspection of the returned device found that it has been opened.The proximal anchor was off the insertion device and the device had been actuated (the torque driver is in the post insertion position) and the distal anchor was screwed onto the distal end of the insertion device.There is no damage to the anchors when viewed under magnification.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the packaging process found that the operator should inspect device/trays for integrity and particulate before loading device into pouch.A review of the anchor material specifications found that the raw material strength requirements and storage requirements are specified and each lot must be accompanied by a material certificate of analysis.The root cause could not be determined since the reported malfunction could not be duplicated during the investigation.Please refer to the instructions for use for precautions and warnings related to the use of the device.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRORAPTOR KNOTLESS SA PEEK
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16537913
Report Number1219602-2023-00377
Device Sequence Number6389373
Product Code MBI
UDI-Device Identifier00885556722527
UDI-Public00885556722527
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K181746
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model Number72205021
Device Catalogue Number72205021
Device Lot Number50931086
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2023
Supplement Date Received by Manufacturer01/16/2024
Initial Report FDA Received Date03/14/2023
Supplement Report FDA Received Date01/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/31/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age16 YR
Patient SexMale
Patient Weight70 KG
-
-