|
Model Number 50000000 |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/22/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
|
Event Description
|
It was reported that the biomed had a arctic sun device that failed the calibration for an error 80 (non-recoverable system error) expected 6 actual 20, biomed ordered a mixing pump and would replace in biomed.Per work order reviewed on 15nov2022, the biomed replaced the mixing pump but device still did not pass calibration expected 6 actual 6.27.Biomed sending device in for evaluation.Per sample evaluation results received on 22feb2023, the root cause of the reported issue was due to a failed tank harness.It was reported that the double bend tube found 1/2 way installed into tank and expanded as was the chiller outlet tube to the tank.It was stated that the tank seals were found to be lifted from the tank and no longer properly sealing.It was also stated that device leaked around chiller tubes due to the seals no longer sealing the tank around the tubes.The hot side connection between the power inlet module and the ac main voltage circuit card shows signs of electrical overstress.It was noted that replaced the tank seals, double bend, and chiller evaporator outlet tubes due to expansion.It was also noted that preventatively replaced the power inlet module and ac main voltage circuit card.
|
|
Event Description
|
It was reported that the biomed had a arctic sun device that failed the calibration for an error 80 (non-recoverable system error) expected 6 actual 20, biomed ordered a mixing pump and would replace in biomed.Per work order reviewed on 15nov2022, the biomed replaced the mixing pump but device still did not pass calibration expected 6 actual 6.27.Biomed sending device in for evaluation.Per sample evaluation results received on 22feb2023, it was reported that the root cause of the reported issue was due to a failed tank harness.It was stated that the double bend tube found 1/2 way installed into tank and expanded as was the chiller outlet tube to the tank.It was also stated that the tank seals were found to be lifted from the tank and no longer properly sealing.It was noted that device leaked around chiller tubes due to the seals no longer sealing the tank around the tubes.The hot side connection between the power inlet module and the ac main voltage circuit card shows signs of electrical overstress.It was noted that replaced the tank seals, double bend, and chiller evaporator outlet tubes due to expansion.It was also noted that the preventatively replaced the power inlet module and ac main voltage circuit card.
|
|
Manufacturer Narrative
|
The reported issue was confirmed.The device was evaluated upon receipt.The hot side connection between the power inlet module and the ac main voltage circuit card shows signs of electrical overstress.Replaced power inlet module and ac main voltage circuit card.The arctic sun 5000 passed all performance testing, calibration and electrical safety tests and is functioning properly and ready for use.Root cause -inadequate verification and validation activities of the crimping process, single pull test did not provide stability of process, evidence was not provided when requested for maintenance of records or crimp tools, no crimp cross-sections provided.The dhr review is not required and the risk review and labeling review is not required as the investigation was confirmed related to supplier issue.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: the actual/suspected device was inspected.
|
|
Search Alerts/Recalls
|
|
|