• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE G; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE G; IMPLANTABLE LEAD Back to Search Results
Model Number 0185
Device Problem Unexpected Therapeutic Results (1631)
Patient Problem Fall (1848)
Event Date 02/22/2023
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient with this system with an implantable cardioverter defibrillator (icd), right ventricular (rv) and right atrial (ra) leads fell to the ground.A shock was delivered but a fault code 1004 related to short circuit was triggered.Also, a code 1006 regarding high voltage was triggered afterwards.External shocks were needed to convert the arrhythmia the patient was experiencing.The icd, ra and rv leads have been explanted at this time.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE G
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
#12, rd. #698
dorado PR 00646
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16551324
MDR Text Key311369129
Report Number2124215-2023-12406
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00802526531194
UDI-Public00802526531194
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P910073/S043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 03/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/23/2009
Device Model Number0185
Device Catalogue Number0185
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/23/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/23/2007
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age63 YR
Patient SexMale
-
-