INSULET CORPORATION OMNIPOD DASH INSULIN MANAGEMENT SYSTEM; PUMP, INFUSION, INSULIN Back to Search Results
Model Number PT-000010
Device Problems Premature Discharge of Battery (1057); Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/14/2023
Event Type  malfunction  
Manufacturer Narrative
A technical assessment of the device design identified the root cause of the thermal event to be the dash pdm charging voltage exceeding the battery specification, defined as overcharging.Field safety corrective action 9056196-10/11/2022-001-c has been initiated by insulet corporation.The device will not be returned for investigation.
 
Event Description
It was reported that the personal diabetes management (pdm) battery was bulging and was not holding a charge.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNIPOD DASH INSULIN MANAGEMENT SYSTEM
Type of Device
PUMP, INFUSION, INSULIN
Manufacturer (Section D)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer (Section G)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer Contact
michael spears
100 nagog park
acton, MA 01720
9786007000
MDR Report Key16581312
MDR Text Key311789460
Report Number3004464228-2023-07515
Device Sequence Number1
Product Code LZG
UDI-Device Identifier10385082000139
UDI-Public(01)10385082000139(11)201111(10)L000311
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K211575
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 03/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberPT-000010
Device Catalogue NumberINT1-D001-MG
Device Lot NumberL000311
Was Device Available for Evaluation? No
Date Manufacturer Received03/14/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/11/2020
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-