ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752200
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/03/2023
Event Type  malfunction  
Event Description
A physician reported that the phacoemulsification tip was broken during cataract surgery using the system when the surgeon was in the sculpting step making the groove, and at one point it was noticed that the tip was not eating and that it was pushing the nucleus.When it was removed from the eye, it was observed that the tip was broken in the sleeve.The tip was reused for this surgery.There was no patient injury.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
A sample was not received at the manufacturing site for evaluation; however, the attached customer photo confirms the reported issue of the doctor withdrew the tip and we noticed that it was broken.The broken portion of the tip was contained within the ultra-sleeve (pink).No lot number was identified with this complaint; therefore, a device history record review could not be conducted.The report of phaco tip broken is confirmed based on the photo attached to parent complaint, however, because a sample was not received and no lot number was provided, the root cause for the customer complaint issue cannot be determined.The exact root cause for this complaint is unknown; therefore, specific action with regards to this complaint cannot be taken.All phaco tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Any nonconformance, such as the broken phaco tip exhibited on the photo attached to the parent complaint, is removed from the lot and scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16642142
MDR Text Key312281536
Report Number1644019-2023-00339
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380657522002
UDI-Public00380657522002
Combination Product (y/n)N
Reporter Country CodeAR
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065752200
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-