It was reported that the right ventricular (rv) lead exhibited loss of capture and the patient almost passed out.Additionally, the impedance was a low within range value.Technical services (ts) provided possible causes of the loss of capture.However, the cause of the loss of capture is unknown.The cardiac resynchronization therapy defibrillator (crt-d) was reprogrammed to left ventricular (lv) pacing only.The lead remains in use.The patient is not in good health.Thus, it was decided no further interventions will be performed.No adverse patient effects were reported.
|