MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG, 1688 CAMERA CONTROL UNIT (CCU); LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG, 1688 CAMERA CONTROL UNIT (CCU); LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 1688010000
Device Problems Overheating of Device (1437); Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/14/2023
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event Description
It was reported that there was a thermal event.
 
Event Description
It was reported that there was a thermal event.
 
Manufacturer Narrative
Alleged failure: per rep, the screen says overheating every 5 cases.Sprayed an air vent.Still overheating.The failure(s) identified in the investigation is consistent with the complaint record.Root cause: defective electrical component failure (altera chip) on the digital board.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, 1688 CAMERA CONTROL UNIT (CCU)
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
benjamin ly
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key16704168
MDR Text Key313005999
Report Number0002936485-2023-00291
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier07613327420081
UDI-Public07613327420081
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K182160
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/07/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1688010000
Device Catalogue Number1688010000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received03/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-