BOSTON SCIENTIFIC CORPORATION RELIANCE 4-FRONT; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number 0672
Device Problems Device Dislodged or Dislocated (2923); High Capture Threshold (3266)
Patient Problem Cardiac Perforation (2513)
Event Date 03/23/2023
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that this right ventricular (rv) lead exhibited high pacing threshold measurements.The physician ordered for chest x-ray and discovered that the lead had an odd position and pericardial perforation was suspected.This rv lead was surgically repositioned and remains in service.No additional adverse patient effects were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The related investigation determined that this lead was associated with a reported perforation with no conclusive evidence of a malfunction; please refer to the description for more information regarding the specific circumstances of this event.
 
Event Description
It was reported that this right ventricular (rv) lead exhibited high pacing threshold measurements.The physician ordered for chest x-ray and discovered that the lead had an odd position and pericardial perforation was suspected.This rv lead was surgically repositioned and remains in service.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELIANCE 4-FRONT
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16709317
MDR Text Key312977669
Report Number2124215-2023-16642
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00802526592829
UDI-Public00802526592829
Combination Product (y/n)Y
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/10/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number0672
Device Catalogue Number0672
Device Lot Number195737
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/26/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/31/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention; Hospitalization;
Patient Age66 YR
Patient SexMale
-
-