Model Number 0672 |
Device Problems
Device Dislodged or Dislocated (2923); High Capture Threshold (3266)
|
Patient Problem
Cardiac Perforation (2513)
|
Event Date 03/23/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that this right ventricular (rv) lead exhibited high pacing threshold measurements.The physician ordered for chest x-ray and discovered that the lead had an odd position and pericardial perforation was suspected.This rv lead was surgically repositioned and remains in service.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The related investigation determined that this lead was associated with a reported perforation with no conclusive evidence of a malfunction; please refer to the description for more information regarding the specific circumstances of this event.
|
|
Event Description
|
It was reported that this right ventricular (rv) lead exhibited high pacing threshold measurements.The physician ordered for chest x-ray and discovered that the lead had an odd position and pericardial perforation was suspected.This rv lead was surgically repositioned and remains in service.No additional adverse patient effects were reported.
|
|