• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RESOLUTION 360 CLIP; HEMOSTATIC METAL CLIP FOR THE GI TRACT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RESOLUTION 360 CLIP; HEMOSTATIC METAL CLIP FOR THE GI TRACT Back to Search Results
Model Number M00521231
Device Problems Entrapment of Device (1212); Human-Device Interface Problem (2949)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/17/2023
Event Type  malfunction  
Manufacturer Narrative
Block h6: imdrf device code a150208 captures the reportable event of clip stuck in scope.
 
Event Description
Note: this report pertains to one of three resolution 360 clips used in the same patient and procedure.It was reported to boston scientific corporation that three resolution 360 clip devices were used during a colonoscopy procedure performed on (b)(6) 2023.During the procedure, the clip dropped after deployment and went up into the endoscope.This was only discovered when the clip fell out of the scope during cleaning.The same problem occurred with the second and third resolution 360 clips.The procedure was completed with another resolution 360 clip.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTION 360 CLIP
Type of Device
HEMOSTATIC METAL CLIP FOR THE GI TRACT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
propark, coyol
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key16713455
MDR Text Key313638132
Report Number3005099803-2023-01736
Device Sequence Number1
Product Code PKL
UDI-Device Identifier08714729875635
UDI-Public08714729875635
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K222503
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00521231
Device Catalogue Number54773
Device Lot Number0030287513
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/17/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/05/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-