• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUICKFLEX MICRO LV LEAD; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUICKFLEX MICRO LV LEAD; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1258T/86
Device Problem Therapy Delivered to Incorrect Body Area (1508)
Patient Problem Discomfort (2330)
Event Type  Injury  
Event Description
It was reported that the patient had ongoing extra cardiac stimulation with bipolar left ventricle lead.Multiple attempts to reprogram did not solve the issue.The left ventricle lead was explanted and replaced.Patient was stable.
 
Manufacturer Narrative
The reported event was extra cardiac stimulation.As received, a complete lead was returned in one piece for analysis.Visual and x-ray inspection did not find any anomalies on the lead.Electrical testing did not find any indication of conductor fractures or internal shorts.S-curve height of the lead was measured within specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKFLEX MICRO LV LEAD
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16716167
MDR Text Key313058227
Report Number2017865-2023-17257
Device Sequence Number1
Product Code OJX
UDI-Device Identifier05414734502610
UDI-Public05414734502610
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/11/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1258T/86
Device Catalogue Number1258T-86
Device Lot NumberA000112965
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/31/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/22/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR; RIGHT ATRIAL LEAD; RIGHT VENTRICLE LEAD
Patient Outcome(s) Required Intervention;
Patient SexFemale
Patient EthnicityNon Hispanic
-
-